Product event summary: the 4fc12 sheath with lot 0010566287 was returned and analyzed.
Visual inspection prior to disassembly and functional testing was performed on the shaft, handle, and dilator.
The inspection identified a shaft kink/twist at 2.
8 approximately inches from the tip.
The steering mechanism and deflection test revealed a noise heard from the handle when the knob was turned with friction.
Dissection of the returned device revealed a kinked pull wire in the barrel of handle area.
In conclusion, the sheath failed the returned product inspection due to a shaft kink.
If information is provided in the future, a supplemental report will be issued.
|