MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97714 |
Device Problems
Failure to Deliver Energy (1211); Migration or Expulsion of Device (1395); Energy Output Problem (1431); Therapy Delivered to Incorrect Body Area (1508); Inappropriate/Inadequate Shock/Stimulation (1574); Patient Device Interaction Problem (4001)
|
Patient Problems
Undesired Nerve Stimulation (1980); Pain (1994); Discomfort (2330); Ambulation Difficulties (2544); Electric Shock (2554)
|
Event Date 07/30/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.Pt stated they had been dealing with their ins for 7 years now.Pt said the ins hadn't helped since it was put in.Pt said they had bad luck with the reps; pt said they thought they were rude and said one rep told pt the ins was not coming out of their body so pt didn't see reps anymore and the ins hadn't been tuned up in about 4 years.Pt also said they get shocks from "it" and the amount of batteries they put into "this" was ridiculous.Pt said they wanted the ins taken out or have a newer model put in so pt could have an mri (which pt said a person with back issues should be able to have).Pt said they had talked to their implanting healthcare provider (hcp), but pt thinks that hcp was only in it for the money and pt was getting ready to look for a lawyer due to the hcp.Pt said the hcp allowed pt's nerve damage to go on so long that pt could barely walk or play with their granddaughter - pt said they felt 90 even though they were 51.Pt said the last time they spoke with that hcp, they told the pt not to worry about the pain.Pt said they had tried seeing many other hcps, but they wouldn't touch the pt because hcp put the device in.Pt mentioned one hcp thought about putting a pain pump in with the ins because the pt had so much pain and ins hadn't helped.Pss offered physician listings, but pt said they just wanted to speak with a rep and see if they could help the pt with their situation.Pss emailed field team in pt's area.The patient mentioned unrelated medical history including pt was on depression meds.
|
|
Event Description
|
The patient reported that the therapy does not help.They are in a lot of pain and can barely walk.They only feel stimulation in the foot (where they need it) if they bend over, otherwise they feel it in their ribs.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a healthcare professional (hcp).The hcp reported that pt had a sudden loss of stimulator c overage.Pt was receiving stimulation more in the upper rib cage which was bothersome, and was not receiving coverage into the areas of their pain.Pt attempted multiple rounds of reprogramming of the implantable neurostimulator (ins).It was reported the stimulation was too high.Imaging revealed the migration of the lead.Pt ins was not mri compatible which the pt wished to have.Implant site was moved from the left buttock to the right buttock since there was pain at the ins site/left buttock area.Lead and ins was replaced.Pt was in stable condition and hcp reported there were no apparent complications.
|
|
Manufacturer Narrative
|
D10: section d information references the main component of the system and other applicable components are the leads.Product id 39565-65 lot# serial# (b)(6) ,implanted: (b)(6) 2015, explanted: (b)(6) 2022, product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the ins (s/n: (b)(6) revealed that the implantable neurostimulator (ins) passed functional testing.Analysis of the lead (s/n: (b)(6) revealed the lead was returned segmented; however electrical testing of the returned segment(s) determined that continuity was complete and there were no electrical shorts between the circuits.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|