Model Number MV-F502927 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that when about 20% of the stent was deployed, the distal part of the fred slipped and the microcatheter moved to proximal.The fred was attempted to be re-sheathed for re-deployment, but dropped off in the microcatheter.The fred was then retrieved along with the microcatheter and another set of fred and a microcatheter were used to continue the procedure, which was completed successfully.There was no impact to the patient.
|
|
Manufacturer Narrative
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.
|
|
Manufacturer Narrative
|
Summary of device evaluation: the reported complaint is unconfirmed as the stent did not detach prematurely at any point during the investigation.The investigation found no other damage or anomaly that would have contributed to the reported complaint.The returned microcatheter was found to be compressed at the distal section; however, there is no indication that this condition resulted in the stent releasing just distal to the hub.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|
|