MAUDE Adverse Event Report: MICROVENTION, INC. FRED 27; FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. FRED 27; FLOW DIVERTER Back to Search Results
Model Number MV-F502927
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2021
Event Type  malfunction  
Event Description
It was reported that when about 20% of the stent was deployed, the distal part of the fred slipped and the microcatheter moved to proximal.The fred was attempted to be re-sheathed for re-deployment, but dropped off in the microcatheter.The fred was then retrieved along with the microcatheter and another set of fred and a microcatheter were used to continue the procedure, which was completed successfully.There was no impact to the patient.
 
Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.
 
Manufacturer Narrative
Summary of device evaluation: the reported complaint is unconfirmed as the stent did not detach prematurely at any point during the investigation.The investigation found no other damage or anomaly that would have contributed to the reported complaint.The returned microcatheter was found to be compressed at the distal section; however, there is no indication that this condition resulted in the stent releasing just distal to the hub.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRED 27
Type of Device
FLOW DIVERTER
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key13232990
MDR Text Key285047617
Report Number2032493-2022-00017
Device Sequence Number1
Product Code OUT
UDI-Device Identifier04987892121968
UDI-Public(01)04987892121968(11)201021(17)230930(10)20102156W
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P180027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberMV-F502927
Device Lot Number20102156W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-