It was reported that during a cryoablation procedure, the balloon catheter was replaced and the case continued.Ablations were performed and the patient became hypotensive.A pericardial effusion was detected on intracardiac echo.The patient was stabilized under anesthesia.The case was aborted while the patient was under general anesthesia.After the case, the patient had additional surgery to drain the effusion.No further patient complications have been reported as a result of this event.
|
Product event summary: the data files were returned and analyzed.The files showed at least six were performed with a returned balloon catheter.Flow fluctuation and a pressure drop was observed on the second application.The inflation phase could not be sustained, and the balloon collapsed.Additional files showed more applications were performed with non-returned balloon catheter without any issue on the date of the event.In conclusion, the clinical issue, pericardial effusion, occurred during the procedure.There is no indication of a relationship of the adverse event to the performance or malfunction of the product.The physical product was not returned.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|