Model Number URF-V3
Device Problems Material Protrusion/Extrusion (2979); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The scope was returned to the service center for evaluation. The customer¿s complaint of ¿ the torn bending section rubber with metal protruding¿ was confirmed. The bending section rubber was found chipped and cracked with fibers sticking out. The scope failed the leak test as leaks were noted on the scope¿s light tube, plastic scope cover and forceps passage. Restriction was noted on the scope¿s control body unit. The scope¿s id chip showed 28 total uses. The cause of the physical damage to the scope cannot be determined at this time. An investigation is ongoing to obtain additional information regarding the reported event. If additional information is received this report will be supplemented accordingly.
 
Event Description
The customer reported that during reprocessing the scope¿s bending section rubber was noted to be torn with metal protruding through. There was no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO VIDEOSCOPE
Type of DeviceURETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13236237
MDR Text Key286435360
Report Number8010047-2022-01312
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/03/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-