Model Number 3ZZ*FX25RECA |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/26/2021 |
Event Type
malfunction
|
Event Description
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, it leaked where the manifold line comes off of the oxygenator.No known impact or consequence to patient.Product was changed out.Procedure was completed successfully.
|
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on january 12, 2022.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - corrected lot number).D9 (device availability - added date returned to manufacturer).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #3: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The affected sample was inspected upon receipt to confirm a damaged arterial pigtail.A representative retention sample was reviewed for damage to the unit, specifically in the area of the arterial sampling pigtail.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|