Model Number HST III SYSTEM (3.8MM)
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2021
Event Type  malfunction  
Event Description
The hospital reported that during a coronary artery bypass procedure using hst iii system (3.8mm).The plastic housing of the heartstring cutter became separated after firing.Procedure was completed with the same device.No injury to pt.
 
Manufacturer Narrative
Trackwise id # (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Manufacturer Narrative
Trackwise # (b)(4).Corrected section: h3 device not eval changed from " device not returned" to "device discarded".Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213/67) the review of the historical data indicates that no other similar complaints was reported for the same lot number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period jan 2020 through dec 2021 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.Device discarded: (4115/ 3221/67) despite request and/ or customer indicated that the device was discarded; however, the actual device involved in the adverse event had been already discarded and thus irretrievably lost for testing.H3 other text : device discarded.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III SYSTEM (3.8MM)
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key13253138
MDR Text Key284712474
Report Number2242352-2022-00064
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700314
UDI-Public00607567700314
Combination Product (y/n)N
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/06/2022
Device Model NumberHST III SYSTEM (3.8MM)
Device Catalogue NumberC-HSK-3038
Device Lot Number25160622
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/31/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/06/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
Patient Weight111 KG
-
-