• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
It was reported that the arctic sun device was not cooling, and a functional check showed that the mixing pump had failed. Per follow-up email received on 04jan2022, the initial reporter stated that the unit was being used on a patient at the time of the incident with no injury. The machine just unhooked when the issue took place. The device was out of service and would be sent in for repair.
 
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13253412
MDR Text Key285046162
Report Number1018233-2021-08805
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 01/13/2022 Patient Sequence Number: 1
-
-