Model Number 8781 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2022 |
Event Type
malfunction
|
Event Description
|
Information was received from a foreign healthcare provider (hcp) via a manufacturer's representative (rep) regarding a patient who was receiving baclofen via an implantable pump.On (b)(6) 2022, it was reported that, during a surgical operation, a malfunction was noted in the pin connector/collet between the pump and spinal catheter segments.It was reported that the connector seemed to be locked.It was unknown if the catheter was inserted into the connector when the malfunction was observed or if the connector was locked out of box.The catheter was replaced and the issue was considered resolved at the time of the report.No surgical intervention occurred or was planned.The patient's status was listed as "alive - no injury".
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a company representative.Regarding the circumstances around which the locked connector was discovered, it was clarified that the event occurred during pump implant.Whether it was determined if the connector was locked inside the sterile kit or if the connector locked after the catheter was inserted was not determined and was still pending an answer from the neurosurgeon.The catheter component had been collected as was to be returned to the manufacturer.
|
|
Manufacturer Narrative
|
H3: the catheter was returned and analysis found that the catheter was not fully seated onto the connector and the collet was locked in place.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|