All three abl90 flex analyzers in the operating room show backlog errors.The field service engineer took samples that were measured in the laboratory and measured them on one of these three abl90 flex analyzers (09r1040n053).After each measurement he had a cal backlog error and k+, na+ and ca2+ were all too low compared to the laboratory.He took out the sensor cassette (sc), cleaned the ref contacts, ran a calibration and still had results that were too low compared to the laboratory.All qc levels were fine.He then took this sc and placed it into another analyzer which measured fine before (09r1040n051).Apparently, the sc carried the errors to the other analyzer.The good sc from abl90, 09r1040n051 also worked fine in abl90, 09r1040n053.Data logs have been sent to radiometer medical for investigation.
|
The customer complains about implausible results for electrolytes and ph on a abl90 flex analyzer.Calibration backlog error is also shown.Original measurement (090r1036n019): ck+: 2,9 & 2,9 mmol/l, cna+: 124 & 128 mmol/l, cca2+: 1,01 0,97 mmol/l, ccl-: 104 & 109 mmol/l, ph: 7,414 & 7,441 mmol/l.Comparison measurements abl90 (two other abl90 flex analyzers): ck+: 4,4 & 4,4 mmol/l, cna+: 139 & 140 mmol/l, cca2+: 1,22 & 1,22 mmol/l, ccl-: 106 & 106 mmol/l, ph: 7,405 & 7,412.Comparison measurement (cobas 8000) units are only an assumption since they are not shown on the printout: k: 4,53 mmol/l, na: 138,1 mmol/l , ca2: 2,35 mmol/l, cl: 104,0 mmol/l.The abl90 flex analyzer (090r1036n019) which this complaint is about, has been moved to the laboratory.A replacement analyzer was set up at the operating room / anesthetic recovery room.Calibration backlog error is also shown for two other abl90 flex analyzers in the operating room.The customer owns more than 20 analyzers and according to the customer only these three show up with errors.There has been no harm to patients.
|