MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Insufficient Cooling (1130); Failure to Pump (1502)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that a functional check showed that the mixing pump had failed. The alert 113 (reduced water temperature control) occurred in the cooling phase. The target temperature was 33. 1c, patient temperature was 36. 6c, water temperature was 31c and not adjusting to reflect the patient temperature. The esophageal probe was in place. The second temperature source was taken and correlating, the arctic sun device was placed in manual mode to check chiller. There was no changes observed in the water temperature until 30 seconds, and nurse would tag for biomed. Nurse doesn¿t think that they have another device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13262853
MDR Text Key284995403
Report Number1018233-2022-00043
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data