Model Number 1427329
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2022
Event Type  malfunction  
Manufacturer Narrative
The device is available for return.The device has not been received.
 
Event Description
The event involved a plum burette set where it was reported that a hair was observed in the cassette.There is no report of patient harm, however there was patient involvement as this was noted during patient infusion.
 
Manufacturer Narrative
Received one used list #142730490, plum 150 ml burette set, clave injection site, (b)(6), 114in; lot #5455899.As received, the set was visually inspected and the hair was confirmed to be inside of the cassette, in the fluid path.No other anomalies were found.The probable cause is due to an error during the gowning process during manufacturing.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUM BURETTE CLAVE SITES 114IN NDEHP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key13290229
MDR Text Key284897582
Report Number9615050-2022-00008
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10887787008804
UDI-Public(01)10887787008804(17)240701(10)5455899
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1427329
Device Catalogue Number142730490
Device Lot Number5455899
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-