MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX52OD PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX52OD PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS Back to Search Results
Model Number 1218-87-352
Device Problem Naturally Worn (2988)
Patient Problems Foreign Body Reaction (1868); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/25/2018
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4). Initial reporter occupation: lawyer. (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Pc was created to captured the left hip. On (b)(6) 2011, the patient had a left total hip arthroplasty to address left hip arthroplasty to address left hip arthritis. Depuy components were used during this procedure. No complications were noted during the procedure. (part/lot page 250 of 302) on (b)(6) 2018, the patient had a revision left total hip arthroplasty, to address failed left total hip arthroplasty, wear of articular bearing surface, and elevated of serum cobalt and chromium. The ceramic femoral head and metal liner were revised. Depuy components were implanted during this procedure along with competitor cement. The surgeon reports that due to excessive acetabular anteversion, the patient has developed elevations of cobalt and chromium. The patient reported experiencing subluxation. Prior to surgery the patient had an aspiration, which was noted to suggest infection, yet a repeat aspiration negated that diagnosis. During the surgery, the surgeon observed black synovial fluid (which was later diagnosed as metallosis). A single screw was left in place as the head had stripped, the remaining screws were removed. The cup was left insitu as the surgeon felt they could correct the midposition with a liner. The surgeon then roughed up the acetabular cup with a saw blade and created several notches in the posterior surface of the new poly liner for competitor cement placement. (sticker sheet page 261 of 302). Doi: (b)(6) 2011 dor: (b)(6), 2018 left hip please see pc-(b)(4) for right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePINNACLE MTL INS NEUT36IDX52OD
Type of DevicePINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13290659
MDR Text Key287529602
Report Number1818910-2022-01183
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/04/2015
Device Model Number1218-87-352
Device Catalogue Number121887352
Device Lot Number3234933
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/19/2022 Patient Sequence Number: 1
Treatment
CORAIL2 NON COL HO SIZE 9; DELTA CER HEAD 12/14 36MM +1.5; PINN CAN BONE SCREW 6.5MMX15MM; PINN CAN BONE SCREW 6.5MMX25MM; PINN SECTOR W/GRIPTION 52MM; PINNACLE MTL INS NEUT36IDX52OD
-
-