BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2022
Event Type  malfunction  
Event Description
It was reported that during set up for a cori assisted surgery, that the console screen would not turn on.They checked all relevant cable connections, attempted to start up the console with only the tablet plugged in.Also tried to connect video via display port rather than with hdmi.It seemed as though it was stuck in the boot process.The orange lights around the console were blinking and never solidified.The surgery was cancelled and rescheduled for (b)(6) 2022.Patient was not under anesthesia and was never in the room.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H3, h6: the real intelligence cori, rob10024, (b)(6), intended for use in treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported event.A functional evaluation was performed.The reported problem was confirmed.The graphics card has failed and will not communicate to the display on any of the 4 ports.In addition, the motherboard also failed and will not allow the system to boot.A functioning graphics card and motherboard were installed and the system booted properly and displayed images.The most likely cause of this event is a graphics card and motherboard failure.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored through complaint investigation and post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13295267
MDR Text Key284972052
Report Number3010266064-2022-00045
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-