Catalog Number 955540
Device Problems Fluid Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Manufacturer Narrative
Oxiris has been temporarily approved for use in the us under emergency use authorization eua200164 with a specific indication to treat patients with covid-19 infection. Phone number: (b)(6). A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. The device was not received for evaluation however, photos of the impacted oxiris set were provided. Visual inspection observed a hole in the access line near the luer connector. The reported condition was verified. The most probable cause identified is an improper handling/connection of the component by the end user. Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with one unit of an oxiris set, an external blood leakage was observed due to a hole on the access line. There was no patient injury or medical intervention associated with this event. Additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOXIRIS
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR 69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13314015
MDR Text Key284183172
Report Number8010182-2022-00002
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 01/21/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number955540
Device Lot Number20I2502
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-