C.R. BARD, INC. (BASD) -3006260740 NAVARRE UNIVERSAL DRAIN; CHRONIC CATHETERS Back to Search Results
Catalog Number NNU12LPT
Device Problems Fracture (1260); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 07/2022).
 
Event Description
It was reported that prior to a chronic catheter placement procedure, the conical side of the catheter was allegedly broken.The procedure was completed using another device.There was no patient contact.
 
Event Description
It was reported that prior to a chronic catheter placement procedure, the conical side of the catheter was allegedly broken.The procedure was completed using another device.There was no patient contact.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 12f navarre drainage catheter locking pigtail segment was returned for evaluation.Gross visual, microscopic visual and functional evaluations were performed on the returned device.The investigation is confirmed for the reported fracture and identified hole in catheter issue as small circumferential split was noted proximal to the distal tip of the catheter and a hole was noted on the catheter.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2022).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE UNIVERSAL DRAIN
Type of Device
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13330622
MDR Text Key284281213
Report Number3006260740-2022-00050
Device Sequence Number1
Product Code GBX
UDI-Device Identifier00801741036934
UDI-Public(01)00801741036934
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K951475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNNU12LPT
Device Lot NumberGFDU0293
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-