As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 07/2022).
|
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 12f navarre drainage catheter locking pigtail segment was returned for evaluation.Gross visual, microscopic visual and functional evaluations were performed on the returned device.The investigation is confirmed for the reported fracture and identified hole in catheter issue as small circumferential split was noted proximal to the distal tip of the catheter and a hole was noted on the catheter.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2022).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|