Manufacturer's investigation conclusion: a direct correlation between heartmate 3 left ventricular assist system (lvas), serial number (b)(6) , and the report of cardiac arrhythmia and interference with a subcutaneous implantable cardioverter defibrillator (icd) could not be conclusively established through this evaluation.The submitted log files collectively contained data from (b)(6) 2021 through (b)(6) 2021 and found that the pump operated at the set speed without issue.The system appeared to be operating as intended, and there were no notable alarms active in the log files.The patient remains ongoing on (b)(6) with no further reported issues at this time.The heartmate 3 left ventricular assist system (lvas) instructions for use (ifu), rev.C, is currently available.Section 1 ¿introduction¿ of this document lists cardiac arrhythmia as an adverse event that may be associated with the use of the heartmate 3 lvas.Additionally, section 6 ¿patient care and management¿ lists arrhythmia as a potential late postimplant complication that may be associated with the use of heartmate 3 lvas.The ifu contains warnings in the introduction, surgical procedures, and patient care and management sections that that the pump may cause interference with icds (implantable cardioverter defibrillators).If electromagnetic interference occurs it may lead to inappropriate icd therapy.The occurrence of electromagnetic interference with icd sensing may require adjustment of device sensitivity and/or repositioning the lead.Prior to implanting an icd or ipm (implantable pacemaker) in an hm3 patient, the device to be implanted should be placed in close proximity to the hm3 pump and the telemetry verified.If the patient received a heartmate 3 and has a previously implanted device that is found to be susceptible to programming interference, the manufacturer recommends replacing the icd device with one that is not prone to programming interference.The safety testing and classification section of the ifu states that if the hm3 lvas does cause interference to another device, the user is encouraged to try to correct the interference by reorienting or relocating the equipment, increase the separation between the equipment, connect the equipment to an outlet on a circuit different from that to which other devices are connected, or contact the manufacturer for assistance.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The implant kit was shipped on 28oct2021.No further information was provided.The manufacturer is closing the file on this event.
|