Model Number CLV-S40PRO |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2021 |
Event Type
malfunction
|
Event Description
|
A user facility reported to olympus that the lamp would not light.The problem, as reported to olympus, occurred during a diagnostic laparoscopic procedure.The intended procedure was completed using the same device.There was no patient injury, associated with the problem, reported to olympus.
|
|
Manufacturer Narrative
|
As the device was not returned to olympus for evaluation, a root cause for the reported event cannot be determined.If additional information becomes available or the device is returned for evaluation, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 7 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon, ¿main lamp does not light up¿, was reproduced and was attributed to faulty converter and igniter.However, they could not identify the cause of the fault.Olympus will continue to monitor field performance for this device.
|
|