MAUDE Adverse Event Report: 3M COMPANY 3M¿ ATTEST¿; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

3M COMPANY 3M¿ ATTEST¿; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS Back to Search Results
Model Number 1243RE
Device Problem Problem with Sterilization (1596)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Event Description
Instrument tay was placed in sterilizer with integrator tab which did not turn.Indicator outside of pan turned.Sterile processing manager reran the integrator in the sterilizer and it still did not turn.Manufacturer response for steam integrator tab, (brand not provided) (per site reporter) equipment failure reported to 3m representative.Sample sent out via 3m provided mailing labels.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3M¿ ATTEST¿
Type of Device
INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS
Manufacturer (Section D)
3M COMPANY
3m center, 2510 conway ave
bldg. 275-5w-06
st. paul MN 55144
MDR Report Key13357695
MDR Text Key284449123
Report Number13357695
Device Sequence Number1
Product Code JOJ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1243RE
Device Catalogue Number1243RE
Device Lot NumberSG092023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/24/2022
Event Location Hospital
Date Report to Manufacturer01/26/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-