TOGO MEDIKIT CO., LTD. SUPERCATH 5; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
Catalog Number SP208-01 |
Device Problems
Fracture (1260); Material Separation (1562)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 12/27/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The device involved in the event was returned.The returned device was examined and the repeatability test (tensile strength testing etc.) was conducted using reserved samples with the same lot number as that involved in the event.Also, the investigation was conducted by reviewing the records of the manufacturing processes of the iv catheter with the same lot number as that involved in the event, and it was confirmed that there were no manufacturing processes that caused or contributed to the event, and there were no manufacturing records of visual inspections that showed the cause of or contribution to the event.Judging from this examination, a possible cause of this fracture is that a patient himself/herself has pulled the device with a force greater than acceptable level.In result, catheter could not endure the tensile strength, then it was broken.Lot#: 21j12s5, 21k02s6 and 21k29s7.
|
|
Event Description
|
On (b)(6) 2021, at a hospital in (b)(6), it was reported that supercath5 safety i.V.Catheter was found to be fractured by the confirmation of the device by a nurse because of a removal of the catheter from a patient's body by the patient during an infusion.The damaged part was confirmed in the patient's body, but the patient is being followed up.There was no reported patient injury as a result of this event.
|
|