Model Number SGC0701
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot. Additionally, a review of the complaint history identified no other incidents reported from this lot. Based on available information and without a device to analyze a cause for the reported crack could not be determined. The reported leak was a cascading event of reported crack. There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
This is filed to report leak. It was reported that during preparation of a steerable guide catheter (sgc), a very small crack in the flush port was observed. The stop cock was closed when the flush port started leaking. The sgc was not used in the patient and the procedure was completed with a new sgc. There was no patient involvement and no there was no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameG4 STEERABLE GUIDING CATHETER
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13358720
MDR Text Key288891145
Report Number2024168-2022-00858
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 01/26/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number10819R206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-