Model Number 190766 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2022 |
Event Type
malfunction
|
Event Description
|
A user facility biomedical technician (biomed) reported to fresenius that melting or a growth was observed coming from valve 108 of a 2008t machine.The machine alarmed with a v108 error message.There was no reported patient involvement and it is unknown what state the machine was in when the damaged valve 108 was observed.Additional information was requested, however a response was not received.No parts were returned to the manufacturer for a physical evaluation.
|
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Manufacturer Narrative
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformance, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
|
Event Description
|
A user facility biomedical technician (biomed) reported to fresenius that melting or a growth was observed coming from valve 108 of a 2008t machine.The machine alarmed with a v108 error message.There was no reported patient involvement and it is unknown what state the machine was in when the damaged valve 108 was observed.Additional information was requested, however a response was not received.No parts were returned to the manufacturer for a physical evaluation.
|
|