Model Number 407449
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2022
Event Type  malfunction  
Event Description
During the atrial fibrillation procedure, air was aspirated from the introducer.The introducer was replaced and the procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
One 8f swartz braided introducer sheath was received for evaluation.The dilator was also received.The sheath passed pressure and aspiration leak testing with no anomalies observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8F
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13367460
MDR Text Key284989459
Report Number3005334138-2022-00040
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205733
UDI-Public05414734205733
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number407449
Device Catalogue Number407449
Device Lot Number8093115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-