U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000E
Device Problem Insufficient Cooling (1130)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/09/2022
Event Type  Injury  
Event Description
It was reported that the patient was placed on arctic sun device post arrest with target temperature of 36c. Patient had been on 9 hours and weighed 100kg. Patient was covered well with four pads. Water temperature has been 5 degrees, trend in middle. Patient temperature could not get below 36. 9c, flow rate was 2. 2 l/m, alert 52 (extended period of cold water). Patient was on propofol and 25 fentanyl. Heater was off vent tubed. Nurse tried small dose of paralytic this morning and did not help in adjusting water temperature. It was noted that the patient was covid positive. Biomed would talk to team about potentials for heat generation such as infection, central fevers, shivering. Device looked good and would do frequent skin checks.
 
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13374862
MDR Text Key285566985
Report Number1018233-2022-00193
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 01/27/2022 Patient Sequence Number: 1
-
-