Model Number G447
Device Problems Device Alarm System (1012); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Manufacturer Narrative
This product remains implanted and has not been returned.As such, physical analysis has not been conducted in our laboratory.As no further information concerning this report is expected our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this patient contact the physician office regarding the cardiac resynchronization therapy defibrillator (crt-d) device beeping.An in office visit was recommended.The physician reported getting an red alert from the home monitor, for high, shock impedance out of range.The shock impedance has been variable since implant, 81 ohms to 127 ohms.During the office visit, the physician programmed the device to a threshold of 150 ohms, and will monitor the patient closely.The device remains implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13376441
MDR Text Key284646528
Report Number2124215-2021-40475
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/03/2023
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number255888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/16/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-