|
Model Number ILS-1000-CS |
Device Problems
Application Program Problem (2880); Connection Problem (2900)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/07/2022 |
Event Type
Injury
|
Event Description
|
According to the reporter during use, there was extended working channel (ewc) and locatable guide (lg) not connected error.Attempted 2 restarts, unplugging all cables and reconnecting, turning off continues guidance system (cgs) manually, replacing all cables and verifying no physical damage to ports and nothing resolved issue.Verified that all items showed connected except ewc and lg ports.Attempted both lg ports on cgs and both ports show as disconnected.The procedure was not completed and was not completed by other means as it was rescheduled.The patient was under anesthesia.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter during use, there was extended working channel (ewc) and locatable guide (lg) not connected error.Attempted 2 restarts, unplugging all cables and reconnecting, turning off continues guidance system (cgs) manually, replacing all cables and verifying no physical damage to ports and nothing resolved issue.Verified that all items showed connected except ewc and lg ports.Attempted both lg ports on cgs and both ports show as disconnected.The procedure was not completed and was not completed by other means as it was rescheduled.The patient was under anesthesia.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, h3, h6 out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the continues guidance system (cgs) was connected to the qualified system and performing a performance test, it revealed that the cgs was able to power up as intended.The cgs was able to read the locatable guide, sewc, and location board cables and pass local registration as noted.The issue was resolved by replacement of the cgs it was reported that the patient procedure was cancelled due to a system issue and the system did not detect one of the system components.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: g3 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the continues guidance system (cgs) device had no physical damage or physical defect.Functionally, the cgs was connected to the qualified system and performing a performance test, it revealed that the cgs was able to power up as intended.The cgs was able to read the locatable guide, extended working channel (sewc), and location board cables and pass local registration as noted.The issue was resolved by replacement of the cgs.It was reported that the patient procedure was cancelled due to a system issue and the system did not detect one of the system components.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: g1(manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email and phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found no notable conditions.Functional testing noted that the rail was measured, a component on the sensor board was shorted, the power was not supplied to the sensor board and the fuses on the transmitter board and sensor board were blown.It was reported that the patient procedure was cancelled or aborted due to a system issue and the system did not detect one of the system components.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|