• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST STYLUS PLUS; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST STYLUS PLUS; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 792400
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional serial numbers included in this report: (b)(4).13 of 14 devices were returned for evaluation.We are awaiting the return of 1 device.Evaluation of 3 devices found normal chuck wear and the turbines needed to be replaced.The devices were repaired and returned to the customers.Evaluation of 9 devices found chuck wear and lack of maintenance.The devices were cleaned of debris and the turbines were replaced.The devices were repaired and returned to the customers.Evaluation of 1 device found chuck wear and lack of maintenance.The device had a non dentsply sirona manufactured cartridge, which is incompatible.The device was repaired and returned to the customer.
 
Event Description
This report summarizes 14 malfunction events where a midwest stylus plus handpiece would not hold burs.No injury resulted in any of the events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST STYLUS PLUS
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key13401371
MDR Text Key289336909
Report Number9614977-2021-00097
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported14
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number792400
Patient Sequence Number1
-
-