SYNTHES GMBH SAGITTAL SAW ATTACHMENT, LONG, FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.224 |
Device Problems
Mechanical Jam (2983); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition of device being frozen and not moving identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to component failure due to wear.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the saw attachment device was frozen and would not move.It was further determined that the moving parts of the device did not move smoothly, the device did not function and the device had component damage.It was observed that the shaft was stiff and the housing was scratched.It was further determined that the device failed pretest for general condition, checking for mechanical free movement, checking general function in running mode and checking the oscillation frequency with frequency meter.It was noted in the service order that the device was making an abnormal noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|