The vendor was sent photos of the tubing; however, they could not provide a root cause for the tubing failure.The device history record review concluded that, based on the date of manufacture the referenced lot(s) was/were produced according to established procedures and all operators involved were trained to the applicable procedures.There were no accompanying non-conformances.The incoming inspection record of the lot of silicone tubing was reviewed and determined to be acceptable, as the id, od and oal were all in specification and the lot did not contain foreign material.Based on the dhr review, evaluation of the sample images, and review of the incoming inspection record, it cannot be concluded that referenced complaint is manufacturing related.The sterilization and lot history records for the stellaris pack lot# x0108 were reviewed and found to be acceptable.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
|
The initial evaluation has been completed.We received one opened se5525wvb pack, lot x0108 wrapped in a plastic bag, in another plastic bag.Visual inspection found oil on the cassette and tubing.The cassette and tubing are loose and tangled, dirty with solution in the cassette and tubing.Visual inspection also found the iv spike cut off from the tubing and was not returned.The complaint states the infusion line blew; the infusion line has a tear.It cannot be determined the cause of the damage.The infusion line will be sent to the vendor.This investigation is ongoing.
|