BAXTER HEALTHCARE CORPORATION PRISMAX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Device Alarm System (1012)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
Injury
|
Manufacturer Narrative
|
The event occurred on an unreported date in (b)(6) 2022.
Initial reporter pager #: (b)(4).
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during continuous renal replacement therapy with an unspecified number of prismax machines on several patients, black-out screens occurred due to a reboot of the machine following a post failure code 1404 alarm being generated.
The reboot usually takes less than one minute; however, the medical staff was not aware of this feature and did not understand the event.
The medical staff pressed the power button resulting in the therapy recovery option no longer being available.
This resulted approximately one (1) unit of blood loss on an unspecified number of patients.
The extracorporeal blood was not returned to the patients and an unknown number of patients required a blood transfusion.
The patient outcomes were not reported.
No additional information is available.
|
|