Model Number 60000000 |
Device Problems
Insufficient Heating (1287); Overfill (2404)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
|
Event Description
|
It was reported that the nurse stated the patient was not rewarming on the arctic sun device.The nurse wanted to know if the device was working appropriately.The patient temperature was 33.6c, target temperature was 37c, water temperature was 30.4c and flow rate was 3.0lpm.Rewarm from 35.5 c and rewarm rate was 0.33c/hour.When mss walked the nurse through alerts and alarms, alert 113 (reduced water temperature control) was noted.Operating hours were 73, system hours were 102 and heater command was 100 percent.Mss explained how to drain 500 cc off the device and placed the device in manual control with water temperature at 40c.Water temperature immediately went to 38c, and nurse disabled manual mode.Now the patient temperature was 33.8c and water temperature was 37c.
|
|
Manufacturer Narrative
|
The reported issue was confirmed.The root cause of the reported issue was isolated to user related.It is known that the device did not meet specifications and whether the device was influenced by the reported failure.The device was in use on a patient.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "1) drain the reservoir.¿ turn control module power off.¿ attach the drain line to the two drain ports on the back of the control module.Place the end of the drain line into a container.The water will passively drain into the container.2) refill the reservoir.¿ connect the fill tube.¿ from the hypothermia therapy screen or the normothermia therapy screen, press the fill reservoir button.¿ the fill reservoir screen will appear.Follow the directions on the screen.¿ add one vial of arctic sun¿ temperature management system cleaning solution to the second liter of sterile water.¿ the filling process will automatically stop when the reservoir is full.Continue to replace the bottles of sterile water until the filling process stops.¿ when the fill reservoir process is complete, the screen will close.Note: the reservoir level sensor requires a conductive fluid to operate properly.Filling the reservoir without using the arctic sun cleaning solution may result in overfilling the reservoir." the device was not returned.
|
|
Event Description
|
It was reported that the nurse stated the patient was not rewarming on the arctic sun device.The nurse wanted to know if the device was working appropriately.The patient temperature was 33.6c, target temperature was 37c, water temperature was 30.4c and flow rate was 3.0lpm.Rewarm from 35.5 c and rewarm rate was 0.33c/hour.When mss walked the nurse through alerts and alarms, alert 113 (reduced water temperature control) was noted.Operating hours were 73, system hours were 102 and heater command was 100 percent.Mss explained how to drain 500 cc off the device and placed the device in manual control with water temperature at 40c.Water temperature immediately went to 38c, and nurse disabled manual mode.Now the patient temperature was 33.8c and water temperature was 37c.
|
|