BECTON DICKINSON UNSPECIFIED BD LUTONIX¿ CATHETER INTRAVASCULAR CATHETER Back to Search Results
Catalog Number UNKNOWN
Device Problems Fracture (1260); Material Puncture/Hole (1504); Air/Gas in Device (4062)
Patient Problems Bruise/Contusion (1754); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2022
Event Type  malfunction  
Manufacturer Narrative
Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured. A catalog and lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured. Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number. Date of birth: unknown. Medical device expiration date: unknown. Initial reporter phone#: (b)(6). A device evaluation is anticipated, but has not yet begun. Upon completion of the investigation, a supplemental report will be filed. Device manufacture date: unknown.
 
Event Description
It was reported unspecified bd lutonix¿ catheter had a hole in the needle hub. The following information was provided by the initial reporter: "when performing the puncture, it was detected presence of air in the syringe in several occasions without achieving the puncture properly. Therefore, it was performed a review of the puncture needle and the syringe, and it was found the needle was broken in its base which represented the air leakage to the syringe. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNSPECIFIED BD LUTONIX¿ CATHETER
Type of DeviceINTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas NJ 88780
MX 88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13441231
MDR Text Key286932971
Report Number2243072-2022-00133
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K050310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 04/04/2022
2 Devices were Involved in the Event: 1   2  
0 Patients were Involved in the Event:
Date FDA Received02/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-