Model Number G158
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Patient-Device Incompatibility (2682)
Patient Problem Electric Shock (2554)
Event Date 03/03/2021
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited oversensing of noise, resulting in a pause less than six seconds.Anti-tachycardia pacing (atp) has been delivered three times for this behavior, with no out of range pace impedance measurements noted.The physician noted this patient was not pace dependent and ordered a chest x-ray.Technical services (ts) discussed the device system with the physician and recommended programming options.Subsequently, device sensitivity was reprogrammed for optimization and a successful defibrillation threshold (dft) test was completed.No additional adverse patient effects were reported.The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13444913
MDR Text Key285006494
Report Number2124215-2022-02446
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public00802526534904
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/04/2017
Device Model NumberG158
Device Catalogue NumberG158
Device Lot Number134833
Was Device Available for Evaluation? No
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/04/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
Patient Age74 YR
Patient SexMale
-
-