• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR MP FP T1 PPS 6X97.5MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR MP FP T1 PPS 6X97.5MM; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-149060
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in the process of being returned to zimmer biomet for investigation.Once the investigation has been completed a follow-up mdr will be submitted report source: (b)(6).
 
Event Description
It was reported that during inspection at the distributorship a team member discovered hair like debris in sterile packaging.No adverse consequences reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The reported event is confirmed.Evaluation of the returned product confirmed foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the provided work instructions during manufacturing.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR MP FP T1 PPS 6X97.5MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13451062
MDR Text Key287825673
Report Number0001825034-2022-00275
Device Sequence Number1
Product Code KWL
UDI-Device Identifier00880304517431
UDI-Public(01)00880304517431(17)310922(10)7109729
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number51-149060
Device Lot Number7109729
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received04/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-