Model Number MMT-1715KL
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem Hypoglycemia (1912)
Event Date 01/31/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump was over delivering.The customer¿s blood glucose value was 65 mg/dl.Customer had been using insulin pump system within 48 hours of reported low blood glucose event.The customer treated low blood glucose value with food.The customer was neither in the emergency room, nor admitted into hospital as a result of low blood glucose.Customer alleging possible insulin pump was over delivery because he was getting too much insulin while in auto mode.No harm requiring medical intervention was reported.The device will be returned for analysis.
 
Manufacturer Narrative
Retainer ring = black.On 01/31/2022 the customer reported low bg (under delivery).Inserted a test p-cap into the retainer ring, and it locked in place properly.Unit passed displacement, rewind, prime/seating, basic occlusion, force sensor, occlusion, sleep current measurement, active current measurement, self, and displacement accuracy tests.No under or over delivery anomalies were noted during testing.Attempt to upload the unit¿s history and trace files using thus was successful.Attempt to upload the patient¿s data using carelink was also successful.The adapt tool confirms that the following related alarms/alerts occurred around the event date: 100=bolus not delivered alarm occurred 01/16/2022 15:42:33, and 01/24/2022 18:54:17.The adapt tool does not list any pump errors that could potentially trigger a critical pump error.The case was cut open.After visual inspection, did not note any signs of previous moisture presence or corrosion inside the battery compartment or on any of the boards, assemblies, and the motor inside the unit.In summary, the customer¿s report of low bg (under delivery) was not confirmed during testing.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key13468973
MDR Text Key285157881
Report Number2032227-2022-109882
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316655
UDI-Public(01)000000763000316655(17)240910
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG5P3PD
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight88 KG
-
-