Manufacturer ref# pr353161 11) corrected data compared with medwatch report mw5106926 b4) 24mar2022.D1) celect ivc filter.D3) cook medical llc.H6) 3165 device embedded in tissue or placque 1260 fracture.Summary of investigational findings: a celect filter was found to be fractured with half a leg embedded in the spine.The filter and the fragments were removed with forceps.Cook was unable to conduct a review of the device history record, as the lot number of the complaint device was not provided for the investigation.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.There are adequate controls in place to ensure that this type of device is manufactured to specifications.Based on the limited provided information it is unknown what caused the filter to fracture.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|