Model Number ORT200
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/31/2019
Event Type  malfunction  
Manufacturer Narrative
An imris customer service engineer (cse) was on-site to resolve the issue.The cse discovered that the roll cylinder failed to keep back pressure on the line and the cause appeared to be a cylinder valve failure.The cse replaced the roll cylinder with a new one and performed drift testing to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
 
Event Description
On (b)(6) 2019, the customer reported that the ort200 table was exhibiting a large movement in the roll function.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT200 OPERATING ROOM TABLE
Type of Device
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
kwaku amoah
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key13476388
MDR Text Key287281329
Report Number3010326005-2020-00004
Device Sequence Number1
Product Code KXJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Remedial Action Inspection
Type of Report Initial
Report Date 12/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORT200
Device Catalogue Number113821-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse