Model Number 0998-00-0800-31
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2022
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the iabp and found that the device was not connected with the cart. The fse provided training to the staff to address the issue. The unit is working within specification. A supplemental report will be submitted upon completion of our investigation. The full event site name is (b)(6) hospital.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a battery that was not charging. It is unknown the circumstances under which the event occurred. It is also unknown if there was a patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE I (AU)
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13480906
MDR Text Key287030668
Report Number2249723-2022-00241
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567109053
UDI-Public10607567109053
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 02/07/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-31
Device Catalogue Number0998-00-0800-31
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-