Model Number 50000000 |
Device Problem
Insufficient Cooling (1130)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/18/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.
Once the investigation is complete a supplemental report will be filed.
The information provided by bd represents all the known information at this time.
Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
The device was not returned.
|
|
Event Description
|
It was reported that the data file showed alert 113 due to the arctic sun device not cooling and a review of the data showed a failed mixing pump.
Per additional information received on (b)(6) 2022, it was stated that the arctic sun device was not used on patient during the reported issue.
|
|
Manufacturer Narrative
|
The investigation is still in progress.
Once the investigation is complete a supplemental report will be filed.
The information provided by bd represents all the known information at this time.
Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
The device was not returned.
|
|
Event Description
|
It was reported that the data file showed alert 113 due to the arctic sun device not cooling and a review of the data showed a failed mixing pump.
Per additional information received on (b)(6) 2022, it was stated that the arctic sun device was not used on patient during the reported issue.
|
|