WILLIAM COOK EUROPE COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Model Number G52914 |
Device Problems
Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Scar Tissue (2060); Perforation of Vessels (2135); Stenosis (2263); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440); Insufficient Information (4580)
|
Event Type
Injury
|
Event Description
|
It is alleged that the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2012, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
|
|
Manufacturer Narrative
|
Occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
|
Event Description
|
Patient is alleging device tilt, vena cava and organ perforation.Per a venogram: "ivus [intravascular ultrasound] catheter was then used to visualize the iliac veins as well as the inferior vena cava which appeared to be scarred down"."a 22mm x 35mm wallstent was positioned in the inferior vena cava just below the level of the renal vein and alongside the filter.This was deployed and postdilated using a 24mm balloon.The inferior vena cava filter was crushed to the side wall with excellent apposition to the inferior vena cava"."initial venogram demonstrated occluded iliac veins and occluded infrarenal ivc with multiple collaterals and paravertebral veins opacification.After recanalization and stenting, there was brisk flow of contrast through the iliac veins and ivc without any opacification of collaterals.Ivus confirmed widely patent stents through the iliocaval segment without any evidence of residual stenosis".
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ/vena cava (vc) perforation, vc occlusion/scarring, tilt.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.(b)(4) in lot.To date, one other complaint for a similar issue has been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Per (computed tomography) ct report, "5 mm right common iliac artery, 12 mm vertebral, 10 mm mesenteric perforation" "tilting with the apex against the ivc wall" "ivc stenosis status post stent placement" "no fracture, or migration" per ct report, "one of the tines from the ivc filter has penetrated the vertebral body of l4." "impression: 1.Persistent thrombosis of the stented left iliac vein, with collateral vessels in the left groin and abdominal wall".3."displacement of the patient's ivc filter, with one of the tines adjacent to the right common iliac artery.No evidence of pseudoaneurysm or dissection secondary to this finding, however, continued follow-up is recommended".
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: thrombosis, stenosis the reported allegations have been further investigated based on the information provided to date.The additional information regarding thrombosis & stenosis does not change the previous investigation results for vena cava occlusion.20 devices in lot.No other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|