Model Number 0998-00-3023-53 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate this unit.
The fse observed 0 vdc voltage¿s on the power supply's output.
The fse suspects problem with power supply assembly.
The fse needs to check with the working one to confirm the same.
The repairs are ongoing.
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a routine check prior to use on a patient, the cs300 intra-aortic balloon pump (iabp) was not powering on both batteries in addition to the alternative current (a/c) mains.
There was no patient involvement, and no adverse event was reported.
|
|