Model Number 0998-00-3023-53
Device Problem Failure to Power Up (1476)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/19/2022
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate this unit. The fse observed 0 vdc voltage¿s on the power supply's output. The fse suspects problem with power supply assembly. The fse needs to check with the working one to confirm the same. The repairs are ongoing. A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check prior to use on a patient, the cs300 intra-aortic balloon pump (iabp) was not powering on both batteries in addition to the alternative current (a/c) mains. There was no patient involvement, and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13517133
MDR Text Key287030841
Report Number2249723-2022-00273
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 02/11/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? No Answer Provided
Date Manufacturer Received01/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-