Additional narrative: additional procode: hwc complainant part is not expected to be returned for manufacturer review/investigation.Manufacturing location: monument manufacturing date: march 11, 2020 expiration date: january 31, 2029 part: 456.322s, 12mm/130 deg ti cann troch fixation nail 170mm-sterile lot: 46p3459 (sterile) production order traveler met all inspection acceptance criteria.Inspection sheet, in-process / inspect dimensional / final met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Sterilization control number (scn) supplied by ethicon (abq) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to fracture that occurred below the nail¿ does not indicate breakage of the or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|