DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/21/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during preventative maintenance, the cardiosave intra-aortic balloon pump (iabp) unit displayed executive processor board # f5.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the cardiosave intra-aortic balloon pump (iabp) unit, but was unable to reproduce the reported issue.However, the fse observed error code f5.To fix the issue, the fse replaced the pcba,exec processor.The fse is currently waiting on the safety disks that was ordered to complete the repair.A supplemental report will be submitted upon completion of our investigation.The full name of the initial reporter is: (b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(4).
|
|
Search Alerts/Recalls
|
|
|