ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR
|
Back to Search Results |
|
Model Number PM3562 |
Device Problems
High impedance (1291); Failure to Interrogate (1332); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/27/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during initial implant procedure, the patient's new implanted pacemaker exhibited high, out of range pacing impedance.It was also found that the lead was not able to completely connect into pacemaker header and was not able to connect to the programmer.The pacemaker was not implanted during procedure on (b)(6) 2022.An alternate pacemaker was installed.There were no patient consequences.
|
|
Event Description
|
It was reported that during initial implant procedure, the patient's new implanted pacemaker exhibited high, out of range pacing impedance.It was also found that the lead was not able to completely connect into pacemaker header and was not able to connect to the programmer.The pacemaker was not implanted during procedure on (b)(6) 2022.An alternate pacemaker was installed.There were no patient consequences.
|
|
Manufacturer Narrative
|
The complaints of failure to interrogate and high pacing impedance were not confirmed.As received, the device was returned with battery voltage above elective replacement indicator level.Telemetry test, pacer lead impedance test, longevity analysis, header and connector analysis was normal with no anomalies found.
|
|