Model Number 24701
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2022
Event Type  malfunction  
Event Description
It was reported that device component detachment occurred. A 2mm x 40mm x 144cm coyote es balloon catheter was advanced for dilatation. However, during the procedure, the nose cone came off. The physician advanced a new balloon with the wire but felt resistance during introduction. The wire was removed as it seemed to be the issue and it was found that the nose cone was on the distal aspect of the wire. No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOYOTE ES
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13528162
MDR Text Key285742564
Report Number2134265-2022-00864
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K093636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 02/14/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number24701
Device Catalogue Number24701
Device Lot Number0028350795
Was Device Available for Evaluation? No
Date Manufacturer Received01/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No