DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/21/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a routine check, the cardiosave intra-aortic balloon pump (iabp) had a safety disk failed.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate the iabp and was able to reproduce the reported issue.To fix the issue, the fse replaced the condensate removal module, blood detect tubing, and male luer fitting, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|