Model Number 0998-00-0800-45
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2021
Event Type  malfunction  
Event Description
It was reported that prior to use it was observed by the customer that the cardiosave intra-aortic balloon pump (iabp) had a beeping sound when the battery was inserted in slot 1.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) evaluated the iabp unit, and found that the buzzer sound was coming from the iabp unit and not the battery.When verifying the logs there was an alarm code #118 observed.The fse replaced the power management board which corrected the issue and then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.It is expected that the suspected faulty power management board will be returned to getinge's national repair center for failure analysis; a supplemental report will be submitted upon completion of failure analysis.Event site address was shortened due to field character limit and should read: (b)(6).
 
Event Description
It was reported that prior to use it was observed by the customer that the cardiosave intra-aortic balloon pump (iabp) had a beeping sound when the battery was inserted in slot 1.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit, and found that the buzzer sound was coming from the iabp unit and not the battery.When verifying the logs there was an alarm code #118 observed.The fse replaced the power management board which corrected the issue and then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.It is expected that the suspected faulty power management board will be returned to getinge's national repair center for failure analysis; a supplemental report will be submitted upon completion of failure analysis.Event site address was shortened due to field character limit and should read: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13544870
MDR Text Key286082718
Report Number2249723-2022-00299
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-45
Device Catalogue Number0998-00-0800-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-