Model Number 0998-00-0800-45 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to use it was observed by the customer that the cardiosave intra-aortic balloon pump (iabp) had a beeping sound when the battery was inserted in slot 1.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) evaluated the iabp unit, and found that the buzzer sound was coming from the iabp unit and not the battery.When verifying the logs there was an alarm code #118 observed.The fse replaced the power management board which corrected the issue and then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.It is expected that the suspected faulty power management board will be returned to getinge's national repair center for failure analysis; a supplemental report will be submitted upon completion of failure analysis.Event site address was shortened due to field character limit and should read: (b)(6).
|
|
Event Description
|
It was reported that prior to use it was observed by the customer that the cardiosave intra-aortic balloon pump (iabp) had a beeping sound when the battery was inserted in slot 1.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit, and found that the buzzer sound was coming from the iabp unit and not the battery.When verifying the logs there was an alarm code #118 observed.The fse replaced the power management board which corrected the issue and then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.It is expected that the suspected faulty power management board will be returned to getinge's national repair center for failure analysis; a supplemental report will be submitted upon completion of failure analysis.Event site address was shortened due to field character limit and should read: (b)(6).
|
|