ABBOTT VASCULAR ARMADA 35 PTA CATHETER PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1050-040
Device Problems Material Rupture (1546); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2022
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a lesion in the superficial femoral artery (sfa) with moderate calcification and moderate tortuosity. A 5 x 40mm armada 35 balloon dilatation catheter (bdc) was advanced to the target lesion; however resistance met with the anatomy. Then during the first inflation the balloon ruptured at 6 atmospheres (atms) due to calcification. The bdc was removed from the patient and a non-abbott bdc was used to complete the procedure. The physician stated that a calcified nodule at the target lesion caused the balloon rupture. There were no adverse patient effects reported. No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot. The investigation determined the reported complaints appear to be related to circumstances of the procedure. There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 35 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13552652
MDR Text Key288281365
Report Number2024168-2022-01672
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 02/17/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/31/2022
Device Model NumberB1050-040
Device Catalogue NumberB1050-040
Device Lot Number80806G2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-