STANMORE IMPLANTS WORLDWIDE UNKNOWN JTS EXTENSION MECHANISM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number UNK_STM |
Device Problem
Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
|
Event Description
|
Distributor requested a booster jts drive unit.When details were requested, the following was reported: "at the extension.The device did not extend.I understand there is a booster device that can be sent (that is a stronger magnet) is that correct?".
|
|
Manufacturer Narrative
|
Upon further investigation the event of the patient¿s jts implant seizing has previously been reported under pi (b)(4) mdr ref 3004105610-2021-00162.This pi (b)(4) mdr ref 3004105610-2022-00020 is a duplicate.Since the same failure for the same device has been reported twice pi (b)(4) mdr ref 3004105610-2022-00020 will be cancelled as a duplicate.
|
|
Event Description
|
Distributor requested a booster jts drive unit.When details were requested, the following was reported: "at the extension, the device did not extend.I understand there is a booster device that can be sent ( that is a stronger magnet) is that correct?".
|
|
Search Alerts/Recalls
|
|
|