• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.7MM & SMALL SCREWS GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.7MM & SMALL SCREWS GAUGE, DEPTH Back to Search Results
Model Number 319.01
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation. Without a lot number the device history records review could not be completed. Product was not returned. Based on the information available, it has been determined that no corrective and/or preventative action is proposed. This complaint will be accounted for and monitored via post market surveillance activities. If additional information is made available, the investigation will be updated as applicable. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the brown composite handles of the small hex screwdrivers are degrading, leaving brown stains and marks on sterilization containers and some have even begun to chip. Hooks on depth gauges seem worn causing difficulty measuring screw lengths. There was no patient consequence. This report involves one (1) depth gauge for 2. 7mm & small screws. This is report 4 of 6 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEPTH GAUGE FOR 2.7MM & SMALL SCREWS
Type of DeviceGAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ 4614
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13562247
MDR Text Key290539765
Report Number2939274-2022-00560
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation
Type of Report Initial
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number319.01
Device Catalogue Number319.01
Was Device Available for Evaluation? No
Date Manufacturer Received01/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-